DIS SOLVED OX Y GEN TEST KIT
CODE 5856
For determining the dissolved oxygen content of water, this test kit uses
the azide modification of the Winkler Method and employs a LaMotte
Direct Reading Titrator in the final titration.
QUANTITY CONTENTS CODE
30 *Manganous Sulfate Solution *4167-G
30 *Alkaline Potassium Iodide Azide *7166-G
30 *Sulfuric Acid, 1:1 *6141WT-G
60 *Sodium Thiosulfate, 0.025N *4169-H
30 Starch Indicator Solution 4170WT-G
1 Direct Reading Titrator, 0 - 10 Range 0377
1 Test Tube, 5-10-12.9-15-20-25 mL,
glass, w/cap
0608
2 Bottles, Water Sampling, 60 mL, glass 0688-DO
1 Graduated Cylinder, 25 mL, plastic 2-2297
*WARNING: Reagents marked with an * are considered to be potential
health hazards. To view or print a Material Safety Data Sheet (MSDS) for
these reagents go to www.lamotte.com. To obtain a printed copy, contact
LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code
number.
NOTE: A Check Standard is needed to perform an “EPA Accepted”
test.
WARNING! This set contains chemicals
that may be harmful if misused. Read
cautions on individual containers
carefully. Not to be used by children
except under adult supervision
Vue de la page 0
1 2 3 4 5

Résumé du contenu

Page 1

DIS SOLVED OX Y GEN TEST KITCODE 5856For determining the dissolved oxygen content of water, this test kit usesthe azide modification of the Winkler

Page 2 - BarrelPlunger Adapter tip

DI RECT READ ING TITRATOR IN STRUC TIONS1. Fill the titration tube to the specified line with the water sample.2. Add the reagents as specified in

Page 3

DIS SOLVED OX Y GEN TEST PRO CE DURECOL LEC TION & TREAT MENT OF THE WA TER SAM PLESteps 1 through 4 below describe proper sampling technique i

Page 4 - LaMOTTE COM PANY

TITRATION1. Fill the test tube (0608) to the 20 mL line with the “fixed” sampleand cap. NOTE: For more precise oxygen measurements, fill graduatedcyli